The calibration and application of a force-measuring apparatus on the seat of a rowing ergometer

Andrew James Murphy, S.T.H. Chee, A.M.J. Bull, A.H. McGregor

Research output: Contribution to journalArticle

10 Citations (Scopus)

Abstract

A calibration procedure was carried out to optimize the measurement accuracy of an application of force transducers to quantify the force magnitude and centre of pressure on the seat of a rowing ergometer. Known forces were applied to the system at specified locations, and the recordings were used to provide a solution in the least-squares sense for transforming the transducers' voltage outputs to meaningful data. After calibration, the system was capable of measuring force to within 2N of the applied load, identifying the position of force application to within 1.1mm, and was more accurate than other apparatus described in the literature. The potential of the instrumentation for examining aspects of rowing technique in elite athletes was also demonstrated.
Original languageEnglish
Pages (from-to)109-116
Number of pages8
JournalProceedings of the Institution of Mechanical Engineers, Part P: Journal of Sports Engineering and Technology
Volume224
Issue number1
DOIs
Publication statusPublished - Mar 2010

    Fingerprint

Keywords

  • rowing
  • calibrating
  • seat force
  • centre of pressure