    Abstract

    The quality of fit of a trans-tibial patellar tendon bearing (PTB) socket may be influenced by consistency in casting, rectification or alignment. For this study two distinctive different data-capturing concepts were tested in relation to prosthetist performance. The hands-on PTB and hands-off ICECAST compact® concept were studied and compared for inter- and intraprosthetist consistency using a specially designed manikin stump model. A purpose designed digitiser was used to scan a selected surface area of the produced models, 5 for each concept, 10 in total. The extent of casting consistency at each of 936 locations per plaster model was calculated and the level of consistency was quantified. This study has shown that by using the manikin model there is a clear indication that the investigated hands-off concept produces more consistent results than the hands-on concept.
    LanguageEnglish
    Pages100-106
    Number of pages6
    JournalProsthetics and Orthotics International
    Volume27
    Issue number2
    DOIs
    Publication statusPublished - Aug 2004

    Fingerprint

    Manikins
    Patellar Ligament
    Hand

    Keywords

    • trans-tibial casting
    • manikin stump model
    • hands-on PTB
    • ICECAST concepts
    • prosthetics

    Cite this

    @article{9c9c456d96d04caf930ce80b4c196401,
    title = "Pilot study: data-capturing consistency of two trans-tibial casting concepts, using a manikin stump model: a comparison between the hands-on PTB and hands-off ICECAST compact™ concepts",
    abstract = "The quality of fit of a trans-tibial patellar tendon bearing (PTB) socket may be influenced by consistency in casting, rectification or alignment. For this study two distinctive different data-capturing concepts were tested in relation to prosthetist performance. The hands-on PTB and hands-off ICECAST compact{\circledR} concept were studied and compared for inter- and intraprosthetist consistency using a specially designed manikin stump model. A purpose designed digitiser was used to scan a selected surface area of the produced models, 5 for each concept, 10 in total. The extent of casting consistency at each of 936 locations per plaster model was calculated and the level of consistency was quantified. This study has shown that by using the manikin model there is a clear indication that the investigated hands-off concept produces more consistent results than the hands-on concept.",
    keywords = "trans-tibial casting, manikin stump model, hands-on PTB, ICECAST concepts, prosthetics",
    author = "A. Buis and A. Blair and P. Convery and S. Sockalingam and B.F. McHugh",
    note = "Also presented at: International Society for Prosthetics and Orthotics, 11th World Congress, 1-6 August 2004, Hong Kong.",
    year = "2004",
    month = "8",
    doi = "10.1080/03093640308726665",
    language = "English",
    volume = "27",
    pages = "100--106",
    journal = "Prosthetics and Orthotics International",
    issn = "0309-3646",
    number = "2",

    }

    TY - JOUR

    T1 - Pilot study

    T2 - Prosthetics and Orthotics International

    AU - Buis, A.

    AU - Blair, A.

    AU - Convery, P.

    AU - Sockalingam, S.

    AU - McHugh, B.F.

    N1 - Also presented at: International Society for Prosthetics and Orthotics, 11th World Congress, 1-6 August 2004, Hong Kong.

    PY - 2004/8

    Y1 - 2004/8

    N2 - The quality of fit of a trans-tibial patellar tendon bearing (PTB) socket may be influenced by consistency in casting, rectification or alignment. For this study two distinctive different data-capturing concepts were tested in relation to prosthetist performance. The hands-on PTB and hands-off ICECAST compact® concept were studied and compared for inter- and intraprosthetist consistency using a specially designed manikin stump model. A purpose designed digitiser was used to scan a selected surface area of the produced models, 5 for each concept, 10 in total. The extent of casting consistency at each of 936 locations per plaster model was calculated and the level of consistency was quantified. This study has shown that by using the manikin model there is a clear indication that the investigated hands-off concept produces more consistent results than the hands-on concept.

    AB - The quality of fit of a trans-tibial patellar tendon bearing (PTB) socket may be influenced by consistency in casting, rectification or alignment. For this study two distinctive different data-capturing concepts were tested in relation to prosthetist performance. The hands-on PTB and hands-off ICECAST compact® concept were studied and compared for inter- and intraprosthetist consistency using a specially designed manikin stump model. A purpose designed digitiser was used to scan a selected surface area of the produced models, 5 for each concept, 10 in total. The extent of casting consistency at each of 936 locations per plaster model was calculated and the level of consistency was quantified. This study has shown that by using the manikin model there is a clear indication that the investigated hands-off concept produces more consistent results than the hands-on concept.

    KW - trans-tibial casting

    KW - manikin stump model

    KW - hands-on PTB

    KW - ICECAST concepts

    KW - prosthetics

    UR - http://www.informaworld.com/smpp/content~content=a780828466&db=all

    UR - http://dx.doi.org/10.1080/03093640308726665

    U2 - 10.1080/03093640308726665

    DO - 10.1080/03093640308726665

    M3 - Article

    VL - 27

    SP - 100

    EP - 106

    JO - Prosthetics and Orthotics International

    JF - Prosthetics and Orthotics International

    SN - 0309-3646

    IS - 2

    ER -