Mechanical testing and modelling of the Universal 2 implant

M. K. Gislason, E. Foster, D. Main, G. Fusiek, P. Niewczas, M. Bransby-Zachary, D. H. Nash

Research output: Contribution to journalArticle

3 Citations (Scopus)
108 Downloads (Pure)

Abstract

Understanding the load mechanics of orthopaedic implants is important to be able to predict their behaviour in-vivo. Much research, both mechanical and clinical, has been carried out on hip and knee implants, but less has been written about the mechanics of wrist implants. In this paper, the load mechanics of the Universal 2 wrist implant have been measured using two types of measuring techniques, strain gauges and Fibre Bragg Grating measurements to measure strains. The results were compared to a finite element model of the implant. The results showed that the computational results were in good agreement with the experimental results. Better understanding of the load mechanics of wrist implants, using models and experimental results can catalyse the development of future generation implants.

Original languageEnglish
Pages (from-to)511-517
Number of pages7
JournalMedical Engineering and Physics
Volume38
Issue number6
Early online date12 Apr 2016
DOIs
Publication statusPublished - 1 Jun 2016

Keywords

  • finite element modelling
  • mechanical testing
  • strain measurements
  • wrist implants

Fingerprint Dive into the research topics of 'Mechanical testing and modelling of the Universal 2 implant'. Together they form a unique fingerprint.