Mechanical testing and modelling of the Universal 2 implant

M. K. Gislason, E. Foster, D. Main, G. Fusiek, P. Niewczas, M. Bransby-Zachary, D. H. Nash

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Understanding the load mechanics of orthopaedic implants is important to be able to predict their behaviour in-vivo. Much research, both mechanical and clinical, has been carried out on hip and knee implants, but less has been written about the mechanics of wrist implants. In this paper, the load mechanics of the Universal 2 wrist implant have been measured using two types of measuring techniques, strain gauges and Fibre Bragg Grating measurements to measure strains. The results were compared to a finite element model of the implant. The results showed that the computational results were in good agreement with the experimental results. Better understanding of the load mechanics of wrist implants, using models and experimental results can catalyse the development of future generation implants.

LanguageEnglish
Pages511-517
Number of pages7
JournalMedical Engineering and Physics
Volume38
Issue number6
Early online date12 Apr 2016
DOIs
Publication statusPublished - 1 Jun 2016

Fingerprint

Mechanical testing
Mechanics
Wrist
Social Responsibility
Orthopedics
Fiber Bragg gratings
Strain gages
Hip
Knee
Theoretical Models
Research

Keywords

  • finite element modelling
  • mechanical testing
  • strain measurements
  • wrist implants

Cite this

@article{4913aba10042425bb45d4aa1b5725a16,
title = "Mechanical testing and modelling of the Universal 2 implant",
abstract = "Understanding the load mechanics of orthopaedic implants is important to be able to predict their behaviour in-vivo. Much research, both mechanical and clinical, has been carried out on hip and knee implants, but less has been written about the mechanics of wrist implants. In this paper, the load mechanics of the Universal 2 wrist implant have been measured using two types of measuring techniques, strain gauges and Fibre Bragg Grating measurements to measure strains. The results were compared to a finite element model of the implant. The results showed that the computational results were in good agreement with the experimental results. Better understanding of the load mechanics of wrist implants, using models and experimental results can catalyse the development of future generation implants.",
keywords = "finite element modelling, mechanical testing, strain measurements, wrist implants",
author = "Gislason, {M. K.} and E. Foster and D. Main and G. Fusiek and P. Niewczas and M. Bransby-Zachary and Nash, {D. H.}",
year = "2016",
month = "6",
day = "1",
doi = "10.1016/j.medengphy.2016.02.008",
language = "English",
volume = "38",
pages = "511--517",
journal = "Medical Engineering and Physics",
issn = "1350-4533",
number = "6",

}

Mechanical testing and modelling of the Universal 2 implant. / Gislason, M. K.; Foster, E.; Main, D.; Fusiek, G.; Niewczas, P.; Bransby-Zachary, M.; Nash, D. H.

In: Medical Engineering and Physics , Vol. 38, No. 6, 01.06.2016, p. 511-517.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Mechanical testing and modelling of the Universal 2 implant

AU - Gislason, M. K.

AU - Foster, E.

AU - Main, D.

AU - Fusiek, G.

AU - Niewczas, P.

AU - Bransby-Zachary, M.

AU - Nash, D. H.

PY - 2016/6/1

Y1 - 2016/6/1

N2 - Understanding the load mechanics of orthopaedic implants is important to be able to predict their behaviour in-vivo. Much research, both mechanical and clinical, has been carried out on hip and knee implants, but less has been written about the mechanics of wrist implants. In this paper, the load mechanics of the Universal 2 wrist implant have been measured using two types of measuring techniques, strain gauges and Fibre Bragg Grating measurements to measure strains. The results were compared to a finite element model of the implant. The results showed that the computational results were in good agreement with the experimental results. Better understanding of the load mechanics of wrist implants, using models and experimental results can catalyse the development of future generation implants.

AB - Understanding the load mechanics of orthopaedic implants is important to be able to predict their behaviour in-vivo. Much research, both mechanical and clinical, has been carried out on hip and knee implants, but less has been written about the mechanics of wrist implants. In this paper, the load mechanics of the Universal 2 wrist implant have been measured using two types of measuring techniques, strain gauges and Fibre Bragg Grating measurements to measure strains. The results were compared to a finite element model of the implant. The results showed that the computational results were in good agreement with the experimental results. Better understanding of the load mechanics of wrist implants, using models and experimental results can catalyse the development of future generation implants.

KW - finite element modelling

KW - mechanical testing

KW - strain measurements

KW - wrist implants

UR - http://www.scopus.com/inward/record.url?scp=84964361614&partnerID=8YFLogxK

U2 - 10.1016/j.medengphy.2016.02.008

DO - 10.1016/j.medengphy.2016.02.008

M3 - Article

VL - 38

SP - 511

EP - 517

JO - Medical Engineering and Physics

T2 - Medical Engineering and Physics

JF - Medical Engineering and Physics

SN - 1350-4533

IS - 6

ER -