Impact of temperature on ophthalmic viscosurgical devices and clinical implications: a pilot study

Caoimhe Henry, Alejandro Thompson Castro, Mohamed Belkessa, Ewan Rycroft, Ioanna M. Syntouka, Andreia Silva, Fiona Sillars, Mónica S.N. Oliveira, David Lockington*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

Purpose:
To evaluate how temperature affects the rheology of common ophthalmic viscosurgical devices (OVDs) and clinical implications.

Setting:
Tennent Institute of Ophthalmology, Glasgow, with Department of Mechanical and Aerospace Engineering, and Advanced Materials Research Laboratory, University of Strathclyde, Glasgow, United Kingdom.

Design:
Laboratory pilot study.

Methods:
The viscous and elastic responses of 3 OVDs (Eyefill-SC/Eyefill-C/Eyefill-HD) were measured using rotational and extensional rheometers at clinically relevant temperatures (5°C, 25°C, 37°C). Thermal properties were evaluated using differential scanning calorimetry and laser-flash analysis.

Results:
The OVDs tested exhibited viscoelastic properties and shear-thinning behavior. Apparent viscosities and relaxation time were higher at lower temperatures. The Eyefill-C and Eyefill-SC exhibited predominantly viscous character at low frequencies with a transition to predominantly elastic behavior at high frequencies. An increase in temperature led to a decrease in relaxation time under shear and extension. At low frequencies, Eyefill-C and Eyefill-SC moduli increase with decreasing temperatures. Eyefill-HD at 25°C and 37°C displays 2 crossover points, with the storage modulus dominating at low and high frequencies indicating a predominantly elastic behavior. Thermal property analysis revealed Eyefill-C had the lowest thermal conductivity.

Conclusions:
This pilot study confirms our clinical experience that OVD properties are affected by low temperatures, with increased viscosities at low shear rates and higher relaxation times. Cold OVD can cause greater resistance to initiation of intraocular lens (IOL) injection system forces (compared with warmer OVD). Excessively forced injection using cold OVD could contribute to inadvertent cannula detachment if incorrectly assembled, or uncontrolled IOL release leading to avoidable injury.
Original languageEnglish
Pages (from-to)1173-1179
Number of pages7
JournalJournal of Cataract and Refractive Surgery
Volume50
Issue number11
Early online date2 Jul 2024
DOIs
Publication statusPublished - 30 Nov 2024

Keywords

  • ophthalmic viscosurgical devices
  • cataract surgery
  • complications
  • ophthalmic viscoelastic
  • rheology
  • temperature

Fingerprint