High degree of accuracy of a novel image free handheld robot for unicondylar knee arthroplasty in a cadaveric study

Jess H. Lonner, Julie R. Smith, Frederic Picard, Brian Hamlin, Philip J. Rowe, Philip E. Riches

Research output: Contribution to journalArticle

37 Citations (Scopus)

Abstract

Surgical robotics has been shown to improve the accuracy of bone preparation and soft tissue balance in unicondylar knee arthroplasty (UKA). However, although extensive data have emerged with regard to a CT scan-based haptically constrained robotic arm, little is known about the accuracy of a newer alternative, an imageless robotic system. We assessed the accuracy of a novel imageless semiautonomous freehand robotic sculpting system in performing bone resection and preparation in UKA using cadaveric specimens. In this controlled study, we compared the planned and final implant placement in 25 cadaveric specimens undergoing UKA using the new tool. A quantitative analysis was performed to determine the translational, angular, and rotational differences between the planned and achieved positions of the implants. The femoral implant rotational mean error was 1.04° to 1.88° and mean translational error was 0.72 to 1.29 mm across the three planes. The tibial implant rotational mean error was 1.48° to 1.98° and the mean translational error was 0.79 to 1.27 mm across the three planes. The image-free robotic sculpting tool achieved accurate implementation of the surgical plan with small errors in implant placement. The next step will be to determine whether accurate implant placement translates into a clinical and functional benefit for the patient.
LanguageEnglish
Pages206-212
Number of pages7
JournalClinical Orthopaedics and Related Research
Volume473
Issue number1
Early online date8 Jul 2014
DOIs
Publication statusPublished - Jan 2015

Fingerprint

Knee Replacement Arthroplasties
Robotics
Bone and Bones
Thigh

Keywords

  • surgical robotics
  • image free robot
  • knee arthroplasty
  • robotic sculpting system

Cite this

@article{2e381a708cce42a09b69b77793196384,
title = "High degree of accuracy of a novel image free handheld robot for unicondylar knee arthroplasty in a cadaveric study",
abstract = "Surgical robotics has been shown to improve the accuracy of bone preparation and soft tissue balance in unicondylar knee arthroplasty (UKA). However, although extensive data have emerged with regard to a CT scan-based haptically constrained robotic arm, little is known about the accuracy of a newer alternative, an imageless robotic system. We assessed the accuracy of a novel imageless semiautonomous freehand robotic sculpting system in performing bone resection and preparation in UKA using cadaveric specimens. In this controlled study, we compared the planned and final implant placement in 25 cadaveric specimens undergoing UKA using the new tool. A quantitative analysis was performed to determine the translational, angular, and rotational differences between the planned and achieved positions of the implants. The femoral implant rotational mean error was 1.04° to 1.88° and mean translational error was 0.72 to 1.29 mm across the three planes. The tibial implant rotational mean error was 1.48° to 1.98° and the mean translational error was 0.79 to 1.27 mm across the three planes. The image-free robotic sculpting tool achieved accurate implementation of the surgical plan with small errors in implant placement. The next step will be to determine whether accurate implant placement translates into a clinical and functional benefit for the patient.",
keywords = "surgical robotics, image free robot, knee arthroplasty, robotic sculpting system",
author = "Lonner, {Jess H.} and Smith, {Julie R.} and Frederic Picard and Brian Hamlin and Rowe, {Philip J.} and Riches, {Philip E.}",
year = "2015",
month = "1",
doi = "10.1007/s11999-014-3764-x",
language = "English",
volume = "473",
pages = "206--212",
journal = "Clinical Orthopaedics and Related Research",
issn = "0009-921X",
number = "1",

}

High degree of accuracy of a novel image free handheld robot for unicondylar knee arthroplasty in a cadaveric study. / Lonner, Jess H.; Smith, Julie R.; Picard, Frederic; Hamlin, Brian; Rowe, Philip J.; Riches, Philip E.

In: Clinical Orthopaedics and Related Research, Vol. 473, No. 1, 01.2015, p. 206-212.

Research output: Contribution to journalArticle

TY - JOUR

T1 - High degree of accuracy of a novel image free handheld robot for unicondylar knee arthroplasty in a cadaveric study

AU - Lonner, Jess H.

AU - Smith, Julie R.

AU - Picard, Frederic

AU - Hamlin, Brian

AU - Rowe, Philip J.

AU - Riches, Philip E.

PY - 2015/1

Y1 - 2015/1

N2 - Surgical robotics has been shown to improve the accuracy of bone preparation and soft tissue balance in unicondylar knee arthroplasty (UKA). However, although extensive data have emerged with regard to a CT scan-based haptically constrained robotic arm, little is known about the accuracy of a newer alternative, an imageless robotic system. We assessed the accuracy of a novel imageless semiautonomous freehand robotic sculpting system in performing bone resection and preparation in UKA using cadaveric specimens. In this controlled study, we compared the planned and final implant placement in 25 cadaveric specimens undergoing UKA using the new tool. A quantitative analysis was performed to determine the translational, angular, and rotational differences between the planned and achieved positions of the implants. The femoral implant rotational mean error was 1.04° to 1.88° and mean translational error was 0.72 to 1.29 mm across the three planes. The tibial implant rotational mean error was 1.48° to 1.98° and the mean translational error was 0.79 to 1.27 mm across the three planes. The image-free robotic sculpting tool achieved accurate implementation of the surgical plan with small errors in implant placement. The next step will be to determine whether accurate implant placement translates into a clinical and functional benefit for the patient.

AB - Surgical robotics has been shown to improve the accuracy of bone preparation and soft tissue balance in unicondylar knee arthroplasty (UKA). However, although extensive data have emerged with regard to a CT scan-based haptically constrained robotic arm, little is known about the accuracy of a newer alternative, an imageless robotic system. We assessed the accuracy of a novel imageless semiautonomous freehand robotic sculpting system in performing bone resection and preparation in UKA using cadaveric specimens. In this controlled study, we compared the planned and final implant placement in 25 cadaveric specimens undergoing UKA using the new tool. A quantitative analysis was performed to determine the translational, angular, and rotational differences between the planned and achieved positions of the implants. The femoral implant rotational mean error was 1.04° to 1.88° and mean translational error was 0.72 to 1.29 mm across the three planes. The tibial implant rotational mean error was 1.48° to 1.98° and the mean translational error was 0.79 to 1.27 mm across the three planes. The image-free robotic sculpting tool achieved accurate implementation of the surgical plan with small errors in implant placement. The next step will be to determine whether accurate implant placement translates into a clinical and functional benefit for the patient.

KW - surgical robotics

KW - image free robot

KW - knee arthroplasty

KW - robotic sculpting system

UR - http://link.springer.com/journal/11999

U2 - 10.1007/s11999-014-3764-x

DO - 10.1007/s11999-014-3764-x

M3 - Article

VL - 473

SP - 206

EP - 212

JO - Clinical Orthopaedics and Related Research

T2 - Clinical Orthopaedics and Related Research

JF - Clinical Orthopaedics and Related Research

SN - 0009-921X

IS - 1

ER -