Design and validation of an ultrasound array optimised for epidural needle guidance

S. Cochran, G. A. Corner, K. J. Kirk, D. I.A. Lines, M. J. Watson

Research output: Chapter in Book/Report/Conference proceedingConference contribution book

1 Citation (Scopus)

Abstract

Frequent use is made of lumbar epidural and combined spinal epidural anaesthesia and analgesia for pain relief during labour and childbirth. However, the conventional use of palpation to identify the position of the relevant interspinous space can be difficult in obese parturients. The safety, low cost and capability for ultrasound to be used in real time at the point of care make it a suitable clinical imaging modality to use as an alternative. With a conventional biomedical ultrasound imaging array, with or without a needle guide, the needle cannot be made to lie exactly in the imaging plane because of the physical geometry and with lengths up to more than 100 mm for lumbar epidural procedures, it is easy for the needle to leave the imaging plane altogether. The work reported here has been directed towards a practical solution to this problem via custom array design. This takes into account the need for ergonomic manipulation by the anaesthetist, insertion of the needle exactly parallel to the ultrasonic beam, and sterility. The array incorporates two separate electroacoustic components with a space for needle insertion in between. Using standard array modules has allowed this novel configuration to be connected to a conventional imaging system with only minimal software modifications. The array is operated with its body parallel to the obese parturient's back and with an optimised clinical procedure. The procedure first involves use of the array to identify the relevant interspinous space, followed by insertion of the needle through the array, which is subsequently held in place to guide the needle tip until contact is established with the spine or can be withdrawn once the needle path is well established.

LanguageEnglish
Title of host publication2007 IEEE Ultrasonics Symposium Proceedings, IUS
Pages2255-2258
Number of pages4
DOIs
Publication statusPublished - 26 Dec 2007
Event2007 IEEE Ultrasonics Symposium, IUS - New York, NY, United States
Duration: 28 Oct 200731 Oct 2007

Conference

Conference2007 IEEE Ultrasonics Symposium, IUS
CountryUnited States
CityNew York, NY
Period28/10/0731/10/07

Fingerprint

needles
insertion
analgesia
human factors engineering
anesthesia
electroacoustics
spine
pain
labor
manipulators
safety
modules
ultrasonics
computer programs
geometry
configurations

Keywords

  • epidural anaesthesia
  • needle guidance
  • ultrasound array

Cite this

Cochran, S., Corner, G. A., Kirk, K. J., Lines, D. I. A., & Watson, M. J. (2007). Design and validation of an ultrasound array optimised for epidural needle guidance. In 2007 IEEE Ultrasonics Symposium Proceedings, IUS (pp. 2255-2258). [4410140] https://doi.org/10.1109/ULTSYM.2007.567
Cochran, S. ; Corner, G. A. ; Kirk, K. J. ; Lines, D. I.A. ; Watson, M. J. / Design and validation of an ultrasound array optimised for epidural needle guidance. 2007 IEEE Ultrasonics Symposium Proceedings, IUS. 2007. pp. 2255-2258
@inproceedings{5132827d673a4ba09e1ca9e0fbb90468,
title = "Design and validation of an ultrasound array optimised for epidural needle guidance",
abstract = "Frequent use is made of lumbar epidural and combined spinal epidural anaesthesia and analgesia for pain relief during labour and childbirth. However, the conventional use of palpation to identify the position of the relevant interspinous space can be difficult in obese parturients. The safety, low cost and capability for ultrasound to be used in real time at the point of care make it a suitable clinical imaging modality to use as an alternative. With a conventional biomedical ultrasound imaging array, with or without a needle guide, the needle cannot be made to lie exactly in the imaging plane because of the physical geometry and with lengths up to more than 100 mm for lumbar epidural procedures, it is easy for the needle to leave the imaging plane altogether. The work reported here has been directed towards a practical solution to this problem via custom array design. This takes into account the need for ergonomic manipulation by the anaesthetist, insertion of the needle exactly parallel to the ultrasonic beam, and sterility. The array incorporates two separate electroacoustic components with a space for needle insertion in between. Using standard array modules has allowed this novel configuration to be connected to a conventional imaging system with only minimal software modifications. The array is operated with its body parallel to the obese parturient's back and with an optimised clinical procedure. The procedure first involves use of the array to identify the relevant interspinous space, followed by insertion of the needle through the array, which is subsequently held in place to guide the needle tip until contact is established with the spine or can be withdrawn once the needle path is well established.",
keywords = "epidural anaesthesia, needle guidance, ultrasound array",
author = "S. Cochran and Corner, {G. A.} and Kirk, {K. J.} and Lines, {D. I.A.} and Watson, {M. J.}",
year = "2007",
month = "12",
day = "26",
doi = "10.1109/ULTSYM.2007.567",
language = "English",
isbn = "1424413834",
pages = "2255--2258",
booktitle = "2007 IEEE Ultrasonics Symposium Proceedings, IUS",

}

Cochran, S, Corner, GA, Kirk, KJ, Lines, DIA & Watson, MJ 2007, Design and validation of an ultrasound array optimised for epidural needle guidance. in 2007 IEEE Ultrasonics Symposium Proceedings, IUS., 4410140, pp. 2255-2258, 2007 IEEE Ultrasonics Symposium, IUS, New York, NY, United States, 28/10/07. https://doi.org/10.1109/ULTSYM.2007.567

Design and validation of an ultrasound array optimised for epidural needle guidance. / Cochran, S.; Corner, G. A.; Kirk, K. J.; Lines, D. I.A.; Watson, M. J.

2007 IEEE Ultrasonics Symposium Proceedings, IUS. 2007. p. 2255-2258 4410140.

Research output: Chapter in Book/Report/Conference proceedingConference contribution book

TY - GEN

T1 - Design and validation of an ultrasound array optimised for epidural needle guidance

AU - Cochran, S.

AU - Corner, G. A.

AU - Kirk, K. J.

AU - Lines, D. I.A.

AU - Watson, M. J.

PY - 2007/12/26

Y1 - 2007/12/26

N2 - Frequent use is made of lumbar epidural and combined spinal epidural anaesthesia and analgesia for pain relief during labour and childbirth. However, the conventional use of palpation to identify the position of the relevant interspinous space can be difficult in obese parturients. The safety, low cost and capability for ultrasound to be used in real time at the point of care make it a suitable clinical imaging modality to use as an alternative. With a conventional biomedical ultrasound imaging array, with or without a needle guide, the needle cannot be made to lie exactly in the imaging plane because of the physical geometry and with lengths up to more than 100 mm for lumbar epidural procedures, it is easy for the needle to leave the imaging plane altogether. The work reported here has been directed towards a practical solution to this problem via custom array design. This takes into account the need for ergonomic manipulation by the anaesthetist, insertion of the needle exactly parallel to the ultrasonic beam, and sterility. The array incorporates two separate electroacoustic components with a space for needle insertion in between. Using standard array modules has allowed this novel configuration to be connected to a conventional imaging system with only minimal software modifications. The array is operated with its body parallel to the obese parturient's back and with an optimised clinical procedure. The procedure first involves use of the array to identify the relevant interspinous space, followed by insertion of the needle through the array, which is subsequently held in place to guide the needle tip until contact is established with the spine or can be withdrawn once the needle path is well established.

AB - Frequent use is made of lumbar epidural and combined spinal epidural anaesthesia and analgesia for pain relief during labour and childbirth. However, the conventional use of palpation to identify the position of the relevant interspinous space can be difficult in obese parturients. The safety, low cost and capability for ultrasound to be used in real time at the point of care make it a suitable clinical imaging modality to use as an alternative. With a conventional biomedical ultrasound imaging array, with or without a needle guide, the needle cannot be made to lie exactly in the imaging plane because of the physical geometry and with lengths up to more than 100 mm for lumbar epidural procedures, it is easy for the needle to leave the imaging plane altogether. The work reported here has been directed towards a practical solution to this problem via custom array design. This takes into account the need for ergonomic manipulation by the anaesthetist, insertion of the needle exactly parallel to the ultrasonic beam, and sterility. The array incorporates two separate electroacoustic components with a space for needle insertion in between. Using standard array modules has allowed this novel configuration to be connected to a conventional imaging system with only minimal software modifications. The array is operated with its body parallel to the obese parturient's back and with an optimised clinical procedure. The procedure first involves use of the array to identify the relevant interspinous space, followed by insertion of the needle through the array, which is subsequently held in place to guide the needle tip until contact is established with the spine or can be withdrawn once the needle path is well established.

KW - epidural anaesthesia

KW - needle guidance

KW - ultrasound array

UR - http://www.scopus.com/inward/record.url?scp=48149089593&partnerID=8YFLogxK

U2 - 10.1109/ULTSYM.2007.567

DO - 10.1109/ULTSYM.2007.567

M3 - Conference contribution book

SN - 1424413834

SN - 9781424413836

SP - 2255

EP - 2258

BT - 2007 IEEE Ultrasonics Symposium Proceedings, IUS

ER -

Cochran S, Corner GA, Kirk KJ, Lines DIA, Watson MJ. Design and validation of an ultrasound array optimised for epidural needle guidance. In 2007 IEEE Ultrasonics Symposium Proceedings, IUS. 2007. p. 2255-2258. 4410140 https://doi.org/10.1109/ULTSYM.2007.567