Comparison of the results of 4 users of a contemporary CADCAM system

T. McGarry, B. McHugh

    Research output: Contribution to journalArticle

    8 Citations (Scopus)

    Abstract

    The objective of this study was to investigate the variation of measurements recorded when four different users of the Tracer CAD system trace a model of known dimensions and volume. This complements a previous study by the same authors Mc Garry and Mc Hugh, (2005) where the accuracy and consistency of a single user was measured. Landmarks were added to indicate proximal, distal, anterior, medial and lateral regions of a specially manufactured cylindrical nylon 6.6 model. Four circumferential lines were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with guaranteed accuracy to one hundredth of a millimetre, and was traced using the Tracer CAD system by four different users. The difference in mean volume between measured results and Tracer CAD scans of differing users ranged to -3%. Individual trace volumes varied by up to -7.85%. In all volumes measured, eleven out of twelve maximum volume percentage differences measured greater than 2 %, and of these, seven results showed maximum volume percentage difference to measure greater than 4%.
    LanguageEnglish
    Pages27-35
    Number of pages8
    JournalProsthetics and Orthotics International
    Volume31
    Issue number1
    DOIs
    Publication statusPublished - 1 Mar 2007

    Fingerprint

    Nylons

    Keywords

    • CAD CAM
    • shape capture
    • accuracy
    • measurement
    • prosthetics
    • orthotics

    Cite this

    @article{4425cd25b72d4984bcb993e4119156d6,
    title = "Comparison of the results of 4 users of a contemporary CADCAM system",
    abstract = "The objective of this study was to investigate the variation of measurements recorded when four different users of the Tracer CAD system trace a model of known dimensions and volume. This complements a previous study by the same authors Mc Garry and Mc Hugh, (2005) where the accuracy and consistency of a single user was measured. Landmarks were added to indicate proximal, distal, anterior, medial and lateral regions of a specially manufactured cylindrical nylon 6.6 model. Four circumferential lines were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with guaranteed accuracy to one hundredth of a millimetre, and was traced using the Tracer CAD system by four different users. The difference in mean volume between measured results and Tracer CAD scans of differing users ranged to -3{\%}. Individual trace volumes varied by up to -7.85{\%}. In all volumes measured, eleven out of twelve maximum volume percentage differences measured greater than 2 {\%}, and of these, seven results showed maximum volume percentage difference to measure greater than 4{\%}.",
    keywords = "CAD CAM, shape capture, accuracy, measurement, prosthetics, orthotics",
    author = "T. McGarry and B. McHugh",
    year = "2007",
    month = "3",
    day = "1",
    doi = "10.1080/03093640600942101",
    language = "English",
    volume = "31",
    pages = "27--35",
    journal = "Prosthetics and Orthotics International",
    issn = "0309-3646",
    number = "1",

    }

    Comparison of the results of 4 users of a contemporary CADCAM system. / McGarry, T.; McHugh, B.

    In: Prosthetics and Orthotics International, Vol. 31, No. 1, 01.03.2007, p. 27-35.

    Research output: Contribution to journalArticle

    TY - JOUR

    T1 - Comparison of the results of 4 users of a contemporary CADCAM system

    AU - McGarry, T.

    AU - McHugh, B.

    PY - 2007/3/1

    Y1 - 2007/3/1

    N2 - The objective of this study was to investigate the variation of measurements recorded when four different users of the Tracer CAD system trace a model of known dimensions and volume. This complements a previous study by the same authors Mc Garry and Mc Hugh, (2005) where the accuracy and consistency of a single user was measured. Landmarks were added to indicate proximal, distal, anterior, medial and lateral regions of a specially manufactured cylindrical nylon 6.6 model. Four circumferential lines were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with guaranteed accuracy to one hundredth of a millimetre, and was traced using the Tracer CAD system by four different users. The difference in mean volume between measured results and Tracer CAD scans of differing users ranged to -3%. Individual trace volumes varied by up to -7.85%. In all volumes measured, eleven out of twelve maximum volume percentage differences measured greater than 2 %, and of these, seven results showed maximum volume percentage difference to measure greater than 4%.

    AB - The objective of this study was to investigate the variation of measurements recorded when four different users of the Tracer CAD system trace a model of known dimensions and volume. This complements a previous study by the same authors Mc Garry and Mc Hugh, (2005) where the accuracy and consistency of a single user was measured. Landmarks were added to indicate proximal, distal, anterior, medial and lateral regions of a specially manufactured cylindrical nylon 6.6 model. Four circumferential lines were added at regular intervals along the length of the cylinder with a view to calculating diameters and volumes relative to these landmarks. The model was measured using a comparator with guaranteed accuracy to one hundredth of a millimetre, and was traced using the Tracer CAD system by four different users. The difference in mean volume between measured results and Tracer CAD scans of differing users ranged to -3%. Individual trace volumes varied by up to -7.85%. In all volumes measured, eleven out of twelve maximum volume percentage differences measured greater than 2 %, and of these, seven results showed maximum volume percentage difference to measure greater than 4%.

    KW - CAD CAM

    KW - shape capture

    KW - accuracy

    KW - measurement

    KW - prosthetics

    KW - orthotics

    U2 - 10.1080/03093640600942101

    DO - 10.1080/03093640600942101

    M3 - Article

    VL - 31

    SP - 27

    EP - 35

    JO - Prosthetics and Orthotics International

    T2 - Prosthetics and Orthotics International

    JF - Prosthetics and Orthotics International

    SN - 0309-3646

    IS - 1

    ER -