Automated pharmaceutical tablet coating layer evaluation of optical coherence tomography images

Daniel Markl, Günther Hannesschläger, Stephan Sacher, Michael Leitner, Johannes G. Khinast, Andreas Buchsbaum

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

Film coating of pharmaceutical tablets is often applied to influence the drug release behaviour. The coating characteristics such as thickness and uniformity are critical quality parameters, which need to be precisely controlled. Optical coherence tomography (OCT) shows not only high potential for off-line quality control of film-coated tablets but also for in-line monitoring of coating processes. However, an in-line quality control tool must be able to determine coating thickness measurements automatically and in real-time. This study proposes an automatic thickness evaluation algorithm for bi-convex tables, which provides about 1000 thickness measurements within 1 s. Beside the segmentation of the coating layer, optical distortions due to refraction of the beam by the air/coating interface are corrected. Moreover, during in-line monitoring the tablets might be in oblique orientation, which needs to be considered in the algorithm design. Experiments were conducted where the tablet was rotated to specified angles. Manual and automatic thickness measurements were compared for varying coating thicknesses, angles of rotations, and beam displacements (i.e. lateral displacement between successive depth scans). The automatic thickness determination algorithm provides highly accurate results up to an angle of rotation of 30. The computation time was reduced to 0.53 s for 700 thickness measurements by introducing feasibility constraints in the algorithm.

LanguageEnglish
Article number035701
Number of pages12
JournalMeasurement Science and Technology
Volume26
Issue number3
DOIs
Publication statusPublished - 2 Feb 2015

Fingerprint

Optical Coherence Tomography
tablets
Optical tomography
Pharmaceuticals
Drug products
Coating
tomography
Thickness measurement
coatings
Coatings
evaluation
Evaluation
Quality Control
quality control
Quality control
Angle
Monitoring
Refraction
Algorithm Design
Oblique

Keywords

  • distortion correction
  • in-line quality control
  • optical coherence tomography
  • pharmaceutical film-coated tablets
  • segmentation

Cite this

Markl, Daniel ; Hannesschläger, Günther ; Sacher, Stephan ; Leitner, Michael ; Khinast, Johannes G. ; Buchsbaum, Andreas. / Automated pharmaceutical tablet coating layer evaluation of optical coherence tomography images. In: Measurement Science and Technology. 2015 ; Vol. 26, No. 3.
@article{fafc0dcfb8414222b185c27850f8d1f0,
title = "Automated pharmaceutical tablet coating layer evaluation of optical coherence tomography images",
abstract = "Film coating of pharmaceutical tablets is often applied to influence the drug release behaviour. The coating characteristics such as thickness and uniformity are critical quality parameters, which need to be precisely controlled. Optical coherence tomography (OCT) shows not only high potential for off-line quality control of film-coated tablets but also for in-line monitoring of coating processes. However, an in-line quality control tool must be able to determine coating thickness measurements automatically and in real-time. This study proposes an automatic thickness evaluation algorithm for bi-convex tables, which provides about 1000 thickness measurements within 1 s. Beside the segmentation of the coating layer, optical distortions due to refraction of the beam by the air/coating interface are corrected. Moreover, during in-line monitoring the tablets might be in oblique orientation, which needs to be considered in the algorithm design. Experiments were conducted where the tablet was rotated to specified angles. Manual and automatic thickness measurements were compared for varying coating thicknesses, angles of rotations, and beam displacements (i.e. lateral displacement between successive depth scans). The automatic thickness determination algorithm provides highly accurate results up to an angle of rotation of 30. The computation time was reduced to 0.53 s for 700 thickness measurements by introducing feasibility constraints in the algorithm.",
keywords = "distortion correction, in-line quality control, optical coherence tomography, pharmaceutical film-coated tablets, segmentation",
author = "Daniel Markl and G{\"u}nther Hannesschl{\"a}ger and Stephan Sacher and Michael Leitner and Khinast, {Johannes G.} and Andreas Buchsbaum",
year = "2015",
month = "2",
day = "2",
doi = "10.1088/0957-0233/26/3/035701",
language = "English",
volume = "26",
journal = "Measurement Science and Technology",
issn = "0957-0233",
number = "3",

}

Automated pharmaceutical tablet coating layer evaluation of optical coherence tomography images. / Markl, Daniel; Hannesschläger, Günther; Sacher, Stephan; Leitner, Michael; Khinast, Johannes G.; Buchsbaum, Andreas.

In: Measurement Science and Technology, Vol. 26, No. 3, 035701, 02.02.2015.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Automated pharmaceutical tablet coating layer evaluation of optical coherence tomography images

AU - Markl, Daniel

AU - Hannesschläger, Günther

AU - Sacher, Stephan

AU - Leitner, Michael

AU - Khinast, Johannes G.

AU - Buchsbaum, Andreas

PY - 2015/2/2

Y1 - 2015/2/2

N2 - Film coating of pharmaceutical tablets is often applied to influence the drug release behaviour. The coating characteristics such as thickness and uniformity are critical quality parameters, which need to be precisely controlled. Optical coherence tomography (OCT) shows not only high potential for off-line quality control of film-coated tablets but also for in-line monitoring of coating processes. However, an in-line quality control tool must be able to determine coating thickness measurements automatically and in real-time. This study proposes an automatic thickness evaluation algorithm for bi-convex tables, which provides about 1000 thickness measurements within 1 s. Beside the segmentation of the coating layer, optical distortions due to refraction of the beam by the air/coating interface are corrected. Moreover, during in-line monitoring the tablets might be in oblique orientation, which needs to be considered in the algorithm design. Experiments were conducted where the tablet was rotated to specified angles. Manual and automatic thickness measurements were compared for varying coating thicknesses, angles of rotations, and beam displacements (i.e. lateral displacement between successive depth scans). The automatic thickness determination algorithm provides highly accurate results up to an angle of rotation of 30. The computation time was reduced to 0.53 s for 700 thickness measurements by introducing feasibility constraints in the algorithm.

AB - Film coating of pharmaceutical tablets is often applied to influence the drug release behaviour. The coating characteristics such as thickness and uniformity are critical quality parameters, which need to be precisely controlled. Optical coherence tomography (OCT) shows not only high potential for off-line quality control of film-coated tablets but also for in-line monitoring of coating processes. However, an in-line quality control tool must be able to determine coating thickness measurements automatically and in real-time. This study proposes an automatic thickness evaluation algorithm for bi-convex tables, which provides about 1000 thickness measurements within 1 s. Beside the segmentation of the coating layer, optical distortions due to refraction of the beam by the air/coating interface are corrected. Moreover, during in-line monitoring the tablets might be in oblique orientation, which needs to be considered in the algorithm design. Experiments were conducted where the tablet was rotated to specified angles. Manual and automatic thickness measurements were compared for varying coating thicknesses, angles of rotations, and beam displacements (i.e. lateral displacement between successive depth scans). The automatic thickness determination algorithm provides highly accurate results up to an angle of rotation of 30. The computation time was reduced to 0.53 s for 700 thickness measurements by introducing feasibility constraints in the algorithm.

KW - distortion correction

KW - in-line quality control

KW - optical coherence tomography

KW - pharmaceutical film-coated tablets

KW - segmentation

UR - http://www.scopus.com/inward/record.url?scp=84923178972&partnerID=8YFLogxK

U2 - 10.1088/0957-0233/26/3/035701

DO - 10.1088/0957-0233/26/3/035701

M3 - Article

VL - 26

JO - Measurement Science and Technology

T2 - Measurement Science and Technology

JF - Measurement Science and Technology

SN - 0957-0233

IS - 3

M1 - 035701

ER -