TY - JOUR
T1 - A custom, high-channel-count data acquisition system for chemical species tomography of aero-jet engine exhaust plumes
AU - Fisher, Edward M. D.
AU - Tsekenis, Stylianos-Alexios
AU - Yang, Yunjie
AU - Chighine , A.
AU - Liu, Cheng
AU - Polydorides , Nick
AU - Wright, P.
AU - Kliment, J.
AU - Ozanyan, K.
AU - Benoy, Thomas
AU - Humphries, Gordon
AU - Wilson, David
AU - Lengden, M.
AU - Johnstone, Walter
AU - McCann, Hugh
N1 - © 2019 IEEE. Personal use of this material is permitted. Permission from IEEE must be obtained for all other uses, in any current or future media, including reprinting/republishing this material for advertising or promotional purposes, creating new collective works, for resale or redistribution to servers or lists, or reuse of any copyrighted component of this work in other works.
PY - 2020/2/29
Y1 - 2020/2/29
N2 - The fiber-laser imaging of gas turbine exhaust species project aims to provide a video-rate imaging (100 frames/s) diagnostic tool for application to the exhaust plumes of the largest civil aero-jet engines. This remit, enabled by chemical species tomography (CST) currently targeting carbon dioxide (CO 2 ), requires system design that facilitates expansion of multiple parameters. Scalability is needed in order to increase imaging speeds and spatial resolutions and extends the system toward other pertinent gases such as the oxides of nitrogen and sulfur and unburnt hydrocarbons. This paper presents a fully scalable, noninvasive instrument for installation in a commercial engine testing facility, technical challenges having been tackled iteratively through bespoke optical and mechanical design, and it specifically presents the high-speed data acquisition (DAQ) system required. Measurement of gas species concentration is implemented by tunable diode laser absorption with wavelength modulation spectroscopy (TDLAS-WMS) using a custom, high-speed 10-40-MS/s/channel 14-bit DAQ. For CO 2 tomography, the system uses six angular projections of 21 beams each. However, the presented DAQ has capacity for 192 fully parallel 10-Hz-3-MHz differential inputs, achieving a best-case signal-to-noise ratio (SNR) of 56.5 dB prior to filtering. A 12 Ethernet-connected digitization nodes based on field-programmable gate array technology with software control are distributed around a 7-m-diameter mounting “ring.” Hence, the high data rates of 8.96-Gb/s per printed circuit board and 107.52 Gb/s for the whole system can be reduced by using local digital lock-in amplifiers. We believe that this DAQ system is unique in both the TDLAS and CST literatures.
AB - The fiber-laser imaging of gas turbine exhaust species project aims to provide a video-rate imaging (100 frames/s) diagnostic tool for application to the exhaust plumes of the largest civil aero-jet engines. This remit, enabled by chemical species tomography (CST) currently targeting carbon dioxide (CO 2 ), requires system design that facilitates expansion of multiple parameters. Scalability is needed in order to increase imaging speeds and spatial resolutions and extends the system toward other pertinent gases such as the oxides of nitrogen and sulfur and unburnt hydrocarbons. This paper presents a fully scalable, noninvasive instrument for installation in a commercial engine testing facility, technical challenges having been tackled iteratively through bespoke optical and mechanical design, and it specifically presents the high-speed data acquisition (DAQ) system required. Measurement of gas species concentration is implemented by tunable diode laser absorption with wavelength modulation spectroscopy (TDLAS-WMS) using a custom, high-speed 10-40-MS/s/channel 14-bit DAQ. For CO 2 tomography, the system uses six angular projections of 21 beams each. However, the presented DAQ has capacity for 192 fully parallel 10-Hz-3-MHz differential inputs, achieving a best-case signal-to-noise ratio (SNR) of 56.5 dB prior to filtering. A 12 Ethernet-connected digitization nodes based on field-programmable gate array technology with software control are distributed around a 7-m-diameter mounting “ring.” Hence, the high data rates of 8.96-Gb/s per printed circuit board and 107.52 Gb/s for the whole system can be reduced by using local digital lock-in amplifiers. We believe that this DAQ system is unique in both the TDLAS and CST literatures.
KW - aero-engine emissions
KW - aerospace and aviation
KW - chemical species tomography (CST)
KW - data acquisition (DAQ) systems
KW - tunable diode laser absorption spectroscopy (TDLAS)
U2 - 10.1109/TIM.2019.2895932
DO - 10.1109/TIM.2019.2895932
M3 - Article
SN - 0018-9456
VL - 69
SP - 549
EP - 558
JO - IEEE Transactions on Instrumentation and Measurement
JF - IEEE Transactions on Instrumentation and Measurement
IS - 2