Description
Experiments performed in order to assess the possibility to collect good solution phase SAXS data using a conventional diamond anvil cell high pressure cell.
Key findings
Trial tests showed infeasibility of the proposed experiment. A high background induced by the standard thickness DAC diamond windows in a combination with a low scattering power of sample solutions didn’t allow distinguishing a signal from the sample in collected images.
Notes
Experiments performed at Diamond light source (beamline I22) during granted beamline commissioning time
Status | Finished |
---|---|
Effective start/end date | 11/10/18 → 13/10/18 |